Collimator design for a clinical brain SPECT/MRI insert

This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging. We assume the stationary SPECT insert will consist of two rings of ∼5x5-cm SiPM-based detectors insensitive to magnetic fields, with 0.8-mm intrinsic resolution. The maximum diameter is 44.5 cm, the minimum diameter is 33 cm to accommodate the patient and MRI receive/transmit coil, and the FOV has a 20 cm diameter.

We have compared eight collimator designs: single-, 2x2-, 3x3- and 5+2½- pinhole, and single-, 2-, 3- and 1+2½-slit slit-slat, where ½-pinholes/slits are shared between two detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a target resolution of 10 mm FWHM at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. For DOI it is assumed that the crystal's first and second half can be differentiated.

Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and trans-axial resolution, while the latter gives better axial resolution. Although the 2x2-pinhole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilization of the detector FOV. The best options are therefore the 5+2½-pinhole and the 1+2½-slit systems, with sensitivities of 4.9*10–4 and 4.0*10–4, respectively. The brain phantom reconstructions with multi-pinhole collimator are superior as compared to slit-slat, especially in terms of symmetry and realistic activity distribution. DOI information reduces artefacts and improves uniformity in geometric phantoms, although the difference is small for the brain phantom. These results favour a multi-pinhole configuration.

Figure 1
figure 1

Top left: trans-axial sections of a reconstructed uniform phantom. Top right: sagittal sections of a reconstructed Defrise phantom. Bottom left: transaxial sections of a reconstructed Derenzo phantom. Bottom right: transaxial (left), coronal (middle) and sagittal (right column) sections of a reconstructed Zubal brain phantom.

Acknowledgements

DS is supported by a Portuguese PhD Fellowship (FCT: SFRH/BD/88093/2012). This work was done as part of the INSERT collaboration (FP7-HEALTH-2012-INNOVATION-1: 305311). UCL/UCLH research is supported by the NIHR BRCs funding scheme.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Salvado, D., Erlandsson, K., Bousse, A. et al. Collimator design for a clinical brain SPECT/MRI insert. EJNMMI Phys 1 (Suppl 1), A21 (2014). https://doi.org/10.1186/2197-7364-1-S1-A21

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-1-S1-A21

Keywords